4
-
14
2 Select the display type.
[OFF]
: Alarm limit will not be displayed.
[Numerical Value] : Alarm limit will be displayed in numeric format.
[Graph Bar]
: Alarm limit will be displayed in bar graph.
Alarm Limit Setup
The alarm for each parameter can be turned ON or OFF, and upper and lower
alarm limit can be set.
WARNING
Set the appropriate upper and lower alarm limit for each
parameter according to the monitoring condition.
If the alarm limit is not set, alarm limit will not be generated
even if the alarm is set [ON]. Pay attention not to miss any
important alarms.
1 Press the
(Menu) key, and select “Alarm Setup” > “Alarm Limit
Setup”.
The “Alarm Limit Setup” window will
be displayed.
This window can also be displayed by placing the jog dial marker on the
(Alarm Limit Setup) icon and pressing the jog dial.
2 Select ON/OFF and set the upper and lower alarm limit for the parameter
alarm.
(1) Select a parameter for setup.
(2) Select [ON] / [OFF] for the alarm.
If [OFF] is selected, alarm will be turned OFF and
(Alarm OFF) icon will
be displayed.
(3) Aligning the highlight and pressing the jog dial will change the numeric color to
yellow enabling selection of value by turning the jog dial. Press the jog dial to
finalize the setup.
(3)
(1)
(2)
Содержание DSL-8001
Страница 1: ......
Страница 2: ......
Страница 34: ...xxxii ...
Страница 50: ...1 8 ...
Страница 56: ...2 6 Setup 2 2 window ...
Страница 88: ...2 38 Reference For details refer to 4 Monitoring Setup Printing Setup ...
Страница 90: ...2 40 ...
Страница 112: ...3 22 ...
Страница 166: ...6 10 ...
Страница 182: ...7 16 ...
Страница 216: ...8 34 ...
Страница 250: ...10 6 ...
Страница 251: ......
Страница 252: ......