Operating the Analyzer
Operator’s Manual
2
–
35
concentration begins to drop steadily, the permeation device must be replaced.
For details on replacing the permeation device, refer to the Hardware Manual.
Val Start Time
The
Val Start Time
parameter sets the hour of the day for the daily validation
to begin. Enter the required value.
Zero Val Tolerance
The
Zero Val Tolerance
parameter is used to set the maximum acceptable
reading when validating with zero gas. To configure the analyzer for zero gas
set the parameter
Val 1 Concentration
or
Val 2 Concentration
to
0.0
. Enter
the required value.
Adjusting Analyzer Reading to Match Specific
Standard(s)
In some instances, the user may wish to adjust the analyzer reading to match
the concentration (or concentrations) of a specific standard (or standards). The
RATA Multiplier
and
RATA Offset
parameters are used to adjust the analyzer
output in the field without affecting the factory calibration. Both parameters are
used when samples from two different concentration standards are available,
whereas only the
RATA Multiplier
parameter is used when a non-zero gas
sample from only one concentration standard is available. If using a zero gas
sample from only one concentration standard, then only the
RATA Offset
parameter should be calculated.
<SET PARAMETER MODE>
Val Perm Rate Rp
0.33
0:Off >0:ng/min
<SET PARAMETER MODE>
Val Start Time
11
Hour of day (0-23)
<SET PARAMETER MODE>
Zero Val Tolerance
60.0
Enter a value (ppmv)
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...