Serial Port Communications
Operator’s Manual
3
–
23
•
Dry Temp (C)
: Measured temperature (0.0 for non-differential
units) of the gas sample from the last scrubber cycle in degrees
Celsius.
•
Ethane
: Sets the mole fraction of ethane in the dry gas mixture used
when calculating the dew point temperature per ISO 18453:2006.
The default value (0.08360) corresponds to the natural gas mixture
NG3.
•
FitResidue
: Value ranging from 0 to 1 indicating how well the
measured spectrum fits the referenced spectrum, where 1 represents
a perfect match.
•
Fitting Ratio
: Ratio of the measured spectrum versus the Reference
1 or Validation spectrum.
•
Fitting Ratio 2
: Ratio of the measured spectrum versus the
Reference 2 spectrum, where a value of 0 indicates that the
Reference 2 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 3
: Ratio of the measured spectrum versus the
Reference 3 spectrum, where a value of 0 indicates that the
Reference 3 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 4
: Ratio of the measured spectrum versus the
Reference 4 spectrum, where a value of 0 indicates that the
Reference 4 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio 5
: Ratio of the measured spectrum versus the
Reference 5 spectrum, where a value of 0 indicates that the
Reference 5 spectrum was not used in the calculation of
concentration.
•
Fitting Ratio Dry
: Ratio of the measured spectrum versus the
Reference Dry spectrum, where a value of 0 indicates that the
Reference Dry spectrum was not used in the calculation of
concentration.
•
Fitting Ratio Dry-1
: Ratio of the measured spectrum versus the
Reference Dry spectrum shifted by 1 index value, where a value of 0
indicates that the Reference Dry spectrum shifted by 1 index value
was not used in the calculation of concentration.
•
General Alarm DO
: Sets the operation of the general alarm relay
digital output when a general fault alarm occurs and allows it to be
reset if needed.
•
grn ConvFactor 07
: Sets a custom conversion factor when the
parameter Concentration Unit = 7 (grains/100 scf) and this value is
greater than 0.0. The default value conversion factor is used when it
is equal to 0.0.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...