Serial Port Communications
Operator’s Manual
3
–
21
•
AI 20 mA Value
: Sets the pipeline pressure (in mbar) corresponding
to a 20 mA current loop input.
•
AI 4 mA Value
: Sets the pipeline pressure (in mbar) corresponding
to a 4 mA current loop input.
•
AI Pressure Input
: Enables or disables the analog input pipeline
pressure capability.
•
AO 20 mA Value
: Sets the concentration (in ppmv) or dew point
temperature (in degrees Celsius or Fahrenheit), depending on
whether dew point temperature calculation and display is enabled
(i.e., the
Calculate Dew Point
parameter set equal to
1
),
corresponding to a 20 mA current loop output.
•
AO 4 mA Value
: Sets the concentration (in ppmv) or dew point
temperature (in degrees Celsius or Fahrenheit), depending on
whether dew point temperature calculation and display is enabled
(i.e., the
Calculate Dew Point
parameter set equal to
1
),
corresponding to a 4 mA current loop output.
•
AO 4-20 mA Test
: Sets the output of the current loop when in
Mode
5
. The value entered represents a percent of scale value where zero
equals 4 mA and full scale equals 20 mA.
•
Baud Rate
: Sets the baud rate for the customer serial port.
•
Calculate Dew Point
: Enables the calculation of the dew point value
and controls where the value will be output.
•
Cancel Val Alarms
: Cancels all validation alarms and flags that are
activated.
•
Carbon Dioxide
: Sets the mole fraction of carbon dioxide in the dry
gas mixture used when calculating the dew point temperature per
ISO 18453:2006. The default value (0.00000) corresponds to the
natural gas mixture NG3
.
Table 3-2 Assignable Alarm
functionality (Continued)
Bit
Hex Value
Functionality
16
0010000
Ramp adjust restart exceeds limit.
17
0020000
Not used.
18
0040000
Not used.
19
0080000
Flow Switch Alarm.
20
0100000
Validation Fail Alarm 1.
21
0200000
Validation Fail Alarm 2.
22
0400000
Not used.
23
0800000
Not used.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...