NS 5.14 Firmware
2
–
12
4900002262 rev. B 3-23-21
Table 2-1
Typical values for parameter setpoints (Continued)
Parameter
Setting
Function
RATA
(Relative Accuracy Test
Audit)
0 or 1
Default: 0
Enables or disables adjustment
factors.
RATA Multiplier
-1.E+06 to 1.E+06
Default 1.0
Slope adjustment factor.
RATA Offset
-1.E+06 to 1.E+06
Default 0.0
Offset adjustment factor.
Update RATA
0 or 1
Default: 0
Updates RATA Multiplier and
Offset to automatically
calculated values.
Peak Tracking
0, 1 or 2
Default: 0
Sets peak tracking capability to
off, on or reset.
Keypad Watchdog
< 5:Off, >=5:Secs
Default = 10
The time-out on the Mode
screen before the analyzer
automatically reverts to Normal
Mode.
Set Time - Hour
0 to 23
Default: 0
Sets the current hour.
Set Time - Minute
0 to 59
Default: 0
Sets the current minute.
Set Time - Day
1 to 31
Default: 1
Sets the present day.
Set Time - Month
1 to 12
Default: 1
Sets the present month.
Set Time - Year
2006 to 2144
Default: 2017
Sets the present year.
General Alarm DO
0, 1, or 2
Default = 0
Sets the alarm mode (latching
or non-latching) and resets the
general alarm once activated.
DO Alarm Setup
4.29E+09
Default = 8192
Sets the functionality of the
secondary digital output.
Low Alarm Setpoint
-1.E+06 to 1.E+06
Default = -10000
Sets the low alarm threshold
for moving average.
High Alarm Setpoint
-1.E+06 to 1.E+06
Default = 10000
Sets the high alarm threshold
for moving average.
AO 4-20 mA Test
0.0 to 100.0
Default: 0.0
Sets the 4-20 mA output to a
percentage of full scale.
4-20 mA Alarm Action
0, 1, 2 or 3
Default: 0
Sets the current loop state
upon alarm condition.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...