Operating the Analyzer
Operator’s Manual
2
–
33
Tolerance
and
Validation Allowance
)
before stopping the autovalidation
sequence and triggering a
Validation Fail Alarm
. Enter the required value.
Val Auto DumpSpectrm
The
Val Auto DumpSpectrm
parameter determines whether a
Mode 6
dump
automatically occurs after each validation measurement. There are two
choices:
0
to turn the automatic data dump during validation off, and
1
to turn
it on.
Validation Allowance
The
Validation Allowance
parameter sets the tolerance (
%) for validation
measurement when
Val 1 Concentration
or
Val 2 Concentration
is set to a
value greater than
0
.
Validation Wait Time
The
Validation Wait Time
parameter sets the amount of time (in seconds)
for the validation gas to purge the system at the beginning of the validation.
Because validation gas may be introduced into the system at various distances
from the analyzer, adjustment of the
Validation Wait Time
parameter is
necessary to optimize the time the validation gas is allowed to purge through
the transport tubing before the analyzer makes a validation measurement.
Optimization of the
Validation Wait Time
parameter ensures an accurate
measurement of the validation gas while minimizing gas consumption.
<SET PARAMETER MODE>
Val Attempts
2
Enter a value
<SET PARAMETER MODE>
Val Auto DumpSpectrm
0
0:Disable 1:Enable
<SET PARAMETER MODE>
Validation Allowance
10.00000
% of val concentra
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...