NS 5.14 Firmware
2
–
44
4900002262 rev. B 3-23-21
Assignable alarm
The functionality of the
Assignable Alarm
is determined by the
DO Alarm
Setup
parameter set in
Mode 2
according to Table 2–3 on page 2–22. For
example, the
Assignable Alarm
can be configured as a
Concentra High
Alarm
or
Concentra Low Alarm
that is triggered when the measured
concentration is above or below, respectively, the level set in
Mode 2
. A high
concentration causes the
Assignable Alarm
relays to be activated and the
message “Concentration High” displays on the LCD screen.
Validating the Analyzer
Validation of the analyzer using an appropriate gas standard is automatically
conducted at each
Val Interval
period at the
Val Start Time
, semi-
automatically when initiated via the
Validation DI
or the
Start Validation
parameter, or manually by accessing
Mode 7
or
Mode 8
.
To validate automatically
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
If necessary, set the current time, date, desired validation hour and
validation interval (see
“To change parameters in Mode 2”
3.
Set the
Daily Validation
parameter to
1
. The analyzer should run
a validation cycle for the number of days and hour set by
Val
Interval
and the
Val Start Time
parameter.
To validate semi‐automatically
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
Initiate a validation cycle by closing the contacts connected to the
Validation DI input or by setting the
Start Validation
parameter
to
1
.
To validate manually
1.
Verify that the validation gas source(s) has been properly installed
and/or connected.
2.
Initiate validation measurements by pressing the
#
key followed by
the
7
key (
Mode 7
) or the
#
key followed by the
8
key (
Mode 8
)
for a dual validation system.
3.
Once validation measurements are complete, press the
#
key
followed by the
1
key (
Mode 1
) to return to Normal Mode and stop
the validation.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...