Serial Port Communications
Operator’s Manual
3
–
5
•
Fit Ratio Dry
: Ratio of the measured spectrum versus the Reference
Dry spectrum, where a value of 0 indicates that the Reference Dry
spectrum was not used in the calculation of concentration.
•
Fit Ratio Dry-1
: Ratio of the measured spectrum versus the
Reference Dry spectrum shifted by 1 index value, where a value of 0
indicates that the Reference Dry spectrum shifted by 1 index value
was not used in the calculation of concentration.
•
Alarm Flags
: Value representing the status of each individual alarm
in the analyzer, as listed in Table 2–3 on page 2–22.
Any alarm messages will be transmitted along with the data string and
will appear in a separate row.
To capture and save data from the serial port
1.
To save the data from the serial port, use the
Transfer/Capture
Text
function and enter the
Filename
to where you would like to
store the captured data.
2.
To stop the capture of the serial data, click on
Transfer/Capture
Text/Stop
.
To read diagnostic data with HyperTerminal
1.
Before entering
Mode 6
, make sure the serial port on the computer
used for serial communication is connected to the analyzer and the
output stream is showing on the screen as described in
2.
To save the data from the serial port, use the
Transfer/Capture
Text
function and enter the
Filename
to where you would like to
store the captured data.
3.
Once capturing is in place, enter
Mode 6
by pressing
#
key followed
by the
6
key.
The number of seconds between each line of data output should
be the
# Spectrum Average
number set in
Mode 2
divided
by 4. The factory default setting of 16 for #
Spectrum Average
results in a line of output each 4 seconds.
+
#
6
<DUMP SPECTRUM MODE>
Index: 0
Cycle: 1 of 10
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...