Operating the Analyzer
Operator’s Manual
2
–
21
Custom Precision
The
Custom Precision
parameter sets the number of viewable digits to the
right of the decimal point. The total number of digits the analyzer can display
at any one time is six. Therefore, when the size of the value plus the Custom
Precision exceeds six, the number of digits to the right of the decimal point will
be reduced accordingly.
Daily Validation
The
Daily Validation
parameter enables or disables the time of day
autovalidation feature. When enabled, an autovalidation cycle is initiated every
‘X’ day (where ‘X’ is defined by Val Interval) at the time of day established by
Val Start Time
. Enter
0
to turn the feature off or
1
to turn the feature on.
Dew Point Method
The
Dew Point Method
parameter sets the type of industry standard dew
point calculation to be performed when Calculate Dew Point is enabled. Enter
0
for ISO 18453:2006,
1
for ASTM 1142-95 Eq. (1),
2
for ASTM 1142-95 Eq.
(2), and
3
for the Arden Buck Method.
The LCD content can vary slightly depending on system
configuration.
<SET PARAMETER MODE>
Concentration Unit
0
0:ppm 1:lbs 2:% 3:mg
<SET PARAMETER MODE>
Custom Precision
2
Enter a value
<SET PARAMETER MODE>
Daily Validation
0
0:Disable 1:Enable
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...