NS 5.14 Firmware
3
–
22
4900002262 rev. B 3-23-21
•
Common Weight
: Ratio of the measured spectrum that did not
match a reference spectrum.
•
Concentration
: Current measured (live) analyte concentration
(process or validation) in selected engineering units.
•
Concentration (ppmv)
: Current measured (live) analyte
concentration (process or validation) in parts per million by volume
(ppmv).
•
Concentration Process
: Measured analyte concentration of last
process reading in selected engineering units.
•
Concentration Unit
: Designates the display units for the measured
concentration.
•
Conc Process ppmv
: Measured analyte concentration of last
process reading in parts per million by volume (ppmv).
•
Conversion Type
: Sets the type of relations (ideal or real gas) used
in the dew point calculation [
0
for ideal gas relations or
1
for real gas
relations (
Z
calculated using Peng-Robinson EOS)].
•
Cross Shift
: The amount shift applied to match the measured
spectrum to the reference spectrum when using Cross Correlation.
•
Current 2F Flag
: Displays the current scrubber protection usage
level: 0 = standard protection (used for non-differential units), 1 =
mid protection, 2 = high protection, 3 = maximum protection (no
usage).
•
Current Midpoint
: Current midpoint that the analyzer is using,
including any peak tracking adjustments.
•
Custom Precision
: Sets the number of viewable digits to the right
of the decimal point.
•
Daily Validation
: Enables or disables the time of day autovalidation
feature.
•
DC Level
: Signal level at the high end of the current ramp.
•
Dew Point
: Measured analyte concentration of last process reading
as a moisture dew point.
•
Dew Point Method
: Sets the industry standard method with which
the dew point temperature is determined when
Calculate Dew
Point
is enabled.
•
DO Alarm Setup
: Bitwise value that sets the functionality of the
Assignable Alarm, as listed in Table 2–3 on page 2–22.
•
Dry DC Level
: Signal level at the high end of the current ramp from
the last scrubber cycle.
•
Dry Pressure (mb)
: Measured pressure (0.0 for non-differential
units) of the gas sample from the last scrubber cycle in millibar (mb).
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...