Operating the Analyzer
Operator’s Manual
2
–
23
Table 2-3 Assignable Alarm
functionality (Continued)
Bit
Value
Functionality
4
0000010
Laser zero is too low.
5
0000020
Laser zero is too high.
6
0000040
Laser current is too low.
7
0000080
Laser current is too high.
8
0000100
Pressure is too low.
9
0000200
Pressure is too high.
10
0000400
Temperature is too low.
11
0000800
Temperature is too high.
12
0001000
Concentration is too low.
13
0002000
Concentration is too high.
14
0004000
Peak tracking restart exceeds limit.
15
0008000
Fitting restart exceeds limit.
16
0010000
Ramp adjust restart exceeds limit.
17
0020000
Not used.
18
0040000
Not used.
19
0080000
Flow Switch Alarm.
20
0100000
Validation Fail Alarm 1.
21
0200000
Validation Fail Alarm 2.
22
0400000
Not used.
23
0800000
Not used.
24
1000000
Delta DC restart exceeds limit.
25
2000000
Delta temperature restart exceeds limit.
26
4000000
Dry pressure restart exceeds limit.
27
8000000
Need to change scrubber.
28
10000000
R2 restart exceeds limit.
29
20000000
R3 restart exceeds limit.
30
40000000
Pressure delta restart exceeds limit.
31
80000000
Low purge rate alarm.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...