NS 5.14 Firmware
2
–
38
4900002262 rev. B 3-23-21
Application examples
To set manual dual validation
Two standards are used in this example, Validation 1 and Validation 2, which
are manually introduced into the analyzer.
Mode 7
and
Mode 8
are used to
run these standards.
1.
Configure the
Operator Parameter
section with the
Operator
Password
set to
0
and one parameter.
a.
Configure Update RATA by setting
Operator Parameter01
to
“225”. Other parameters can be added, if desired.
2.
Introduce Validation 1 and press
#
followed by the
7
key (to enter
Mode 7
) to allow Validation 1 to run for the desired amount of time.
3.
Introduce Validation 2 and press
#
followed by the
8
key (to enter
Mode 8
) to allow validation to run for the desired amount of time.
4.
Press the
#
key followed by the
2
key (to enter
Mode 2
) and then
the
*
key.
5.
View the newly calculated RATA values, and if desired, set
Update
RATA
to
1
accept the changes. Leave the setting at
0
to reject the
changes and start over.
6.
Re-run steps 2 and 3.
7.
Press the
#
key followed by the
9
key (to enter
Mode 9
) to verify
the validation results and confirm that new RATA values are
operating correctly.
8.
Press the
#
key
followed by the
1
key (to enter
Mode 1
) to return
to Normal Mode.
To set semi‐automatic single or dual validation
In this scenario, either one or two standards (Validation 1 and/or Validation 2)
are used, which are automatically introduced and controlled by the analyzer.
Typically, the validation is initiated locally by the user or while the user is
present by initiating the
Start Validation
parameter or the Start Validation
digital input.
1.
Configure the Operator Parameter section with the
Operator
Password
set to
0
and two parameters.
a.
Configure
Start Validation
by setting Operator Parameter01 to
71
.
b.
Configure
Update RATA
by setting Operator Parameter02 to
225
. Other parameters can be added, if desired.
3.
Press the
#
key
followed by the
2
key (to enter
Mode 2
) and then
press the
*
key.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...