NS 5.14 Firmware
3
–
20
4900002262 rev. B 3-23-21
Modbus accessible parameter definitions
For definitions of the Modbus accessible parameters shown in Table 3–2 below.
For more detailed information, refer to
•
% ConvFactor 02
: Sets a custom conversion factor when the
parameter Concentration Unit = 2 (%) and this value is greater than
0.0. When it is equal to 0.0, the default conversion factor is used.
•
2 Way Com Port
: Sets the port that allows two-way
communications, including Modbus and the diagnostic protocol.
•
4-20 mA Alarm Action
: Determines the current loop state upon an
alarm condition.
•
4-20 mA Input Value
: The current value of the 4-20 mA input for
pipeline pressure in mA’s.
•
4-20 mA Output Value
: The current value of the 4-20 mA output
for concentration in mA’s.
•
4-20 mA Val Action
: Sets the operation mode of the 4-20 mA
current loop during validation cycles.
•
Alarm Flags
: Bitwise value identifying the status of each individual
alarm in the analyzer, as listed in Table 3–2 below.
Table 3–2 Assignable Alarm
functionality
Bit
Hex Value
Functionality
0
0000001
Any alarm active or happened in the past.
1
0000002
Any alarm active (general fault).
2
0000004
Laser power is too low.
3
0000008
Laser power is too high.
4
0000010
Laser zero is too low.
5
0000020
Laser zero is too high.
6
0000040
Laser current is too low.
7
0000080
Laser current is too high.
8
0000100
Pressure is too low.
9
0000200
Pressure is too high.
10
0000400
Temperature is too low.
11
0000800
Temperature is too high.
12
0001000
Concentration is too low.
13
0002000
Concentration is too high.
14
0004000
Peak tracking restart exceeds limit.
15
0008000
Fitting restart exceeds limit.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...