Operating the Analyzer
Operator’s Manual
2
–
31
Start Validation
The
Start
Validation
parameter initiates the validation cycle. After the cycle
begins, this parameter automatically reverts to 0
.
Temperature Unit
The
Temperature Unit
parameter designates the display units for the
measured cell temperature. There are two choices:
0
for degrees Celsius and
1
for Fahrenheit. The default value is the standard unit of measurement in the
region the analyzer is being used.
Update RATA
The
Update RATA
parameter is used to update the
RATA Multiplier
and
RATA Offset
parameters with the latest automatically calculated values. Each
time an automatic, semi-automatic, manual
Mode 7
or manual
Mode 8
validation is completed, a new
RATA Multiplier
and
RATA Offset
will be
calculated. This parameter displays the current
RATA Multiplier
and
RATA
Offset
in the left column of the display as defined in their respective parameter
descriptions. In the right column, the newly calculated values are displayed. To
accept and use the newly calculated values, enter
1
. These values will be
invoked when pressing the MODE button while exiting
Mode 2
.
Refer to the section called
“Adjusting Analyzer Reading to
on page 2-35 section for more
information.
<SET PARAMETER MODE>
Start Validation
1
1:Start
<SET PARAMETER MODE>
Temperature Unit
0
0:C 1:F
<SET PARAMETER MODE>
Mult: 1.00 New: 1.00
Ofst: 0.00 New: 0.00
0
1: Update RATA
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...