NS 5.14 Firmware
2
–
32
4900002262 rev. B 3-23-21
Val 1 Concentration
The
Val 1 Concentration
parameter sets the concentration value of validation
gas supply #1. The analyzer can be configured for a zero gas by setting this
parameter to 0.0, and then setting the
Zero Val Tolerance
to the maximum
acceptable reading. Otherwise, set this parameter to the concentration value
of the validation gas supply and set
Validation Allowance
to the allowable
variation range (±%).
Val 2 Concentration
The
Val 2 Concentration
parameter sets the concentration value of validation
gas supply #2. The analyzer can be configured for a zero gas by setting this
parameter to 0.0, and then setting the
Zero Val Tolerance
to the maximum
acceptable reading. Otherwise, set this parameter to the concentration value
of the validation gas supply and set
Validation Allowance
to the allowable
variation range (±%).
Val Attempts
The
Val Attempts
parameter sets the maximum number of times the analyzer
attempts to measure the validation gas within the set tolerances (see
Zero Val
When procuring a gas standard, make sure the background gas is
that specified or a mix that closely resembles the contents of the
process stream. Have the gas standard certified to better than the
specified precision of the analyzer, if possible.
When procuring a gas standard, make sure the background gas is
that specified or a mix that closely resembles the contents of the
process stream. Have the gas standard certified to better than the
specified precision of the analyzer, if possible.
This parameter only applies to permeation type validation.
<SET PARAMETER MODE>
Val 1 Concentration
60.0
0:ZeroGas >0:ppmvVal
<SET PARAMETER MODE>
Val 2 Concentration
280.000
0:ZeroGas >0:ppmvVal
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...