NS 5.14 Firmware
3
–
30
4900002262 rev. B 3-23-21
•
Validation Allowance
: Sets the tolerance (±%) for validation
measurements when Val 1 Concentration or Val 2 Concentration is set
to a value greater than 0.
•
Validation Wait Time
: Sets the amount of time (in seconds) for the
validation gas to purge the system at the beginning of the validation.
•
Val Perm Constant Kp
: Sets the system constant for the perm
tube.
•
Val Perm Rate Rp
: Sets the calibrated permeation rate for the perm
tube.
•
Val Purge Period
: Sets the number of seconds upon validation
initiation that a differential analyzer will purge the system with
validation gas before starting a dry cycle.
•
Val Start Time
: Sets the hour of the day for the daily autovalidation
to begin.
•
Val Time
: Time of last validation.
•
Wet Pressure (mb)
: Current measured (live) wet pressure reading
of the gas sample in millibar (mb).
•
Wet Temp (C)
: Current measured (live) temperature of the gas
sample in degrees Celsius.
•
Zero Val Tolerance
: Sets the maximum acceptable reading (in
ppmv) when validating with zero gas (Val 1 Concentration = 0 or Val
2 Concentration = 0).
•
Zero Level
: Signal level when the laser is turned off.
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...