NS 5.14 Firmware
2
–
6
4900002262 rev. B 3-23-21
Mode 3: not used
Mode 4: (System Diagnostic parameters)
Mode 4
displays system diagnostic data. These values may be useful when
troubleshooting the system. Press the
#
key followed by the
4
key.
The diagnostic parameters displayed are:
•
Temp:
Shows the temperature in the measurement cell when normal
sample gas is flowing through it.
•
Pres:
Shows the pressure in the measurement cell when normal
sample gas is flowing through it.
•
DC:
Shows the magnitude of the DC laser power in the measurement
cell when normal sample gas is flowing through it. Acceptable values
are between 800 and 3300. A number below or above this range will
trigger a
Laser Power too Low
or
Laser Power too High
,
respectively (see
on page 2-42) indicating that either the
optics need to be cleaned or there is an alignment problem.
•
Zero:
Shows the detector signal value when the laser is turned off.
It should be in the range of -50 to +50. Outside of this range, a
Laser
Zero Low Alarm
or
Laser Zero High Alarm
•
Fit:
The measure of “goodness of fit” for the last measurement point.
•
Mid:
The laser current set point after adjustment by the peak-
tracking software.
Mode 5: (4-20 mA Test Mode)
Mode 5
is used to turn on the 4-20 mA current loop (at the current set with
the
4-20 mA Test
parameter) for test and calibration purposes. Press the
#
key followed by the
5
key.
While in this mode the analyzer suspends measurement until you
return to
Mode 1, Mode 6, Mode 7 or Mode 8
.
+
#
4
Temp: 50.6 C
Pres: 1103.2 mb
DC: 2672
Zero: 12
Fit: 0.98
Mid: 60.24
Содержание SpectraSensors SS2000e
Страница 2: ......
Страница 4: ......
Страница 8: ...NS 5 14 Firmware iv 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...NS 5 14 Firmware vi 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 12: ...NS 5 14 Firmware viii 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 100: ...NS 5 14 Firmware 4 10 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 106: ...NS 5 14 Firmware A 6 4900002262 rev B 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ...NS 5 14 Firmware Index 6 4900002262 rev A 3 23 21 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 113: ......
Страница 114: ...www spectrasensors com contact 4900002262 Rev B...