Commissioning
Proline Cubemass C 300
84
Hauser
Parameter
Prerequisite
Description
Selection / User
entry / User
interface
Factory setting
Pressure value
In the
Pressure compensation
parameter, the
Fixed value
option is selected.
Enter process pressure to be
used for pressure correction.
Positive floating-
point number
1.01325 bar
External pressure
In the
Pressure compensation
parameter, the
External value
option or the
Current input
1…n
option is selected.
Shows the external process
pressure value.
–
*
Visibility depends on order options or device settings
10.4.4 Displaying the I/O configuration
The
I/O configuration
submenu guides the user systematically through all the parameters
in which the configuration of the I/O modules is displayed.
Navigation
"Setup" menu → I/O configuration
‣
I/O configuration
I/O module 1 to n terminal numbers
I/O module 1 to n information
I/O module 1 to n type
Apply I/O configuration
I/O alteration code
Parameter overview with brief description
Parameter
Description
User interface / Selection /
User entry
Factory setting
I/O module 1 to n terminal numbers Shows the terminal numbers used by the I/O
module.
• Not used
• 26-27 (I/O 1)
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
I/O module 1 to n information
Shows information of the plugged I/O
module.
• Not plugged
• Invalid
• Not configurable
• Configurable
• HART
–
I/O module 1 to n type
Shows the I/O module type.
• Off
• Current output
• Current input
• Status input
• Pulse/frequency/switch
output
• Double pulse output
• Relay output
Off