Proline Cubemass C 300
Diagnostics and troubleshooting
Hauser
163
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
948
Oscillation damping too
high
Check process conditions
S
Warning
984
Condensation risk
1. Decrease ambient temperature
2. Increase medium temperature
S
Warning
1)
Diagnostic behavior can be changed.
12.8 Pending diagnostic events
The
Diagnostics
menu allows the user to view the current diagnostic event and the
previous diagnostic event separately.
To call up the measures to rectify a diagnostic event:
• Via "FieldCare" operating tool → 157
• Via "DeviceCare" operating tool → 157
Other pending diagnostic events can be displayed in the
Diagnostic list
submenu
Navigation
"Diagnostics" menu
Diagnostics
Actual diagnostics
Previous diagnostics
Operating time from restart
Operating time
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface
Actual diagnostics
A diagnostic event has occurred.
Shows the current occured diagnostic
event along with its diagnostic
information.
If two or more messages occur
simultaneously, the message with
the highest priority is shown on
the display.
Symbol for diagnostic
behavior, diagnostic code
and short message.
Previous diagnostics
Two diagnostic events have already
occurred.
Shows the diagnostic event that
occurred prior to the current diagnostic
event along with its diagnostic
information.
Symbol for diagnostic
behavior, diagnostic code
and short message.