Commissioning
Proline Cubemass C 300
128
Hauser
Status input 1 to n simulation
Input signal level 1 to n
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign simulation process variable
–
Select a process variable for
the simulation process that is
activated.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Temperature
• Concentration
• Time period signal
frequency (TPS)
Off
Process variable value
A process variable is selected
in the
Assign simulation
process variable
parameter
Enter the simulation value for
the selected process variable.
Depends on the
process variable
selected
0
Current output 1 to n simulation
–
Switch the simulation of the
current output on and off.
• Off
• On
Off
Current output value
In the
Current output 1 to n
simulation
parameter, the
On
option is selected.
Enter the current value for
simulation.
3.59 to 22.5 mA
3.59 mA
Frequency output 1 to n simulation
In the
Operating mode
parameter, the
Frequency
option is selected.
Switch the simulation of the
frequency output on and off.
• Off
• On
Off
Frequency output 1 to n value
In the
Frequency simulation
1 to n
parameter, the
On
option is selected.
Enter the frequency value for
the simulation.
0.0 to 12
500.0 Hz
0.0 Hz
Pulse output simulation 1 to n
In the
Operating mode
parameter, the
Pulse
option is
selected.
Set and switch off the pulse
output simulation.
For
Fixed value
option:
Pulse width
parameter
(→ 92) defines the
pulse width of the pulses
output.
• Off
• Fixed value
• Down-counting
value
Off
Pulse value 1 to n
In the
Pulse output
simulation 1 to n
parameter,
the
Down-counting value
option is selected.
Enter the number of pulses for
simulation.
0 to 65
535
0
Switch output simulation 1 to n
In the
Operating mode
parameter, the
Switch
option
is selected.
Switch the simulation of the
switch output on and off.
• Off
• On
Off
Switch state 1 to n
–
Select the status of the status
output for the simulation.
• Open
• Closed
Open