Proline Cubemass C 300
Technical data
Hauser
187
Interface/protocol
• Via digital communication:
HART protocol
• Via service interface
• CDI-RJ45 service interface
• WLAN interface
Plain text display
With information on cause and remedial measures
Web browser
Plain text display
With information on cause and remedial measures
Light emitting diodes (LED)
Status information
Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
Diagnostic information via light emitting diodes → 150
Low flow cut off
The switch points for low flow cut off are user-selectable.
Galvanic isolation
The outputs are galvanically isolated:
• from the power supply
• from one another
• from the potential equalization (PE) terminal
Protocol-specific data
Manufacturer ID
0x11
Device type ID
0x3B
HART protocol revision
7
Device description files
(DTM, DD
)
Information and files under:
HART load
Min. 250 Ω
System integration
Information on system integration → 71.
• Measured variables via HART protocol
• Burst Mode functionality
16.5 Power supply
Terminal assignment
Supply voltage
Order code for
"Power supply"
Terminal voltage
Frequency range
Option
D
DC 24 V
±20%
–
Option
E
AC 100 to 240 V –15 to +10%
50/60 Hz