Diagnostics and troubleshooting
Proline Cubemass C 300
162
Hauser
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
599
Custody transfer
logbook full
1. Deactivate custody transfer mode
2. Clear custody transfer logbook (all
30 entries)
3. Activate custody transfer mode
F
Warning
Diagnostic of process
803
Loop current 1 faulty
1. Check wiring
2. Change I/O module
F
Alarm
830
Ambient temperature
too high
Reduce ambient temp. around the
sensor housing
S
Warning
831
Ambient temperature
too low
Increase ambient temp. around the
sensor housing
S
Warning
832
Electronics temperature
too high
Reduce ambient temperature
S
Warning
833
Electronics temperature
too low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
842
Process value below
limit
1. Decrease process value
2. Check application
3. Check sensor
S
Warning
862
Partly filled pipe
1. Check for gas in process
2. Adjust detection limits
S
Warning
882
Input signal faulty
1. Check input signal parameterization
2. Check external device
3. Check process conditions
F
Alarm
910
Tubes not oscillating
1. If available: Check connection cable
between sensor and transmitter
2. Check or replace sensor electronic
module (ISEM)
3. Check sensor
F
Alarm
912
Medium
inhomogeneous
1. Check process cond.
2. Increase system pressure
S
Warning
913
Medium unsuitable
1. Check process conditions
2. Check electronic modules or sensor
S
Warning
915
Viscosity ouf of
specification
1. Avoid 2-phase flow
2. Increase system pressure
3. Verify viscosity and density are
within range
4. Check process conditions
S
Warning
941
API/ASTM temperature
out of specificat.
1. Check process temperature with
selected API/ASTM commodity
group
2. Check API/ASTM-related parameters
S
Warning
942
API/ASTM density out
of specification
1. Check process density with selected
API/ASTM commodity group
2. Check API/ASTM-related parameters
S
Warning
943
API pressure out of
specification
1. Check process pressure with selected
API commodity group
2. Check API related parameters
S
Warning
944
Monitoring failed
Check process conditions for Heartbeat
Monitoring
S
Warning