Proline Cubemass C 300
Operation
Hauser
133
11
Operation
11.1 Reading off the device locking status
Device active write protection:
Locking status
parameter
Operation → Locking status
Function scope of the "Locking status" parameter
Options
Description
None
The access authorization displayed in the
Access status
parameter applies
→ 55. Only appears on local display.
Hardware locked
The DIP switch for hardware locking is activated on the PCB board. This locks write
access to the parameters (e.g. via local display or operating tool) → 131.
SIL locked
The SIL mode is enabled. This locks write access to the parameters (e.g. via local
display or operating tool).
Temporarily locked
Write access to the parameters is temporarily locked on account of internal
processes running in the device (e.g. data upload/download, reset, etc.). Once the
internal processing has been completed, the parameters can be changed once
again.
11.2 Adjusting the operating language
Detailed information:
• To configure the operating language → 77
• For information on the operating languages supported by the measuring device
11.3 Configuring the display
Detailed information:
• On the basic settings for the local display → 102
• On the advanced settings for the local display → 116
11.4 Reading measured values
With the
Measured values
submenu, it is possible to read all the measured values.
Navigation
"Diagnostics" menu → Measured values
‣
Measured values
‣
Measured variables
‣
Input values
‣
Output values
‣
Totalizer