Proline Cubemass C 300
Commissioning
Hauser
101
Navigation
"Setup" menu → Double pulse output
‣
Double pulse output
Signal mode
Master terminal number
Assign pulse output
Measuring mode
Value per pulse
Pulse width
Failure mode
Invert output signal
Parameter overview with brief description
Parameter
Description
Selection / User interface /
User entry
Factory setting
Signal mode
Select the signal mode for the double pulse
output.
• Passive
• Active
• Passive NE
Passive
Master terminal number
Shows the terminal numbers used by the
master of the double pulse output module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Assign pulse output
Select process variable for pulse output.
• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Target mass flow
• Carrier mass flow
• Target volume flow
• Carrier volume flow
• Target corrected volume
flow
• Carrier corrected volume
flow
Off
Measuring mode
Select measuring mode for pulse output.
• Forward flow
• Forward/Reverse flow
• Reverse flow
• Reverse flow compensation
Forward flow
Value per pulse
Enter measured value at which a pulse is
output.
Signed floating-point number
Depends on country and
nominal diameter
Pulse width
Define time width of the output pulse.
0.5 to 2
000 ms
0.5 ms