Proline Cubemass C 300
Diagnostics and troubleshooting
Hauser
167
Info number
Info name
I1635
Reset to delivery settings
I1639
Max. switch cycles number reached
I1643
Custody transfer logbook cleared
I1649
Hardware write protection activated
I1650
Hardware write protection deactivated
I1651
Custody transfer parameter changed
I1712
New flash file received
I1725
Sensor electronic module (ISEM) changed
I1726
Configuration backup failed
12.11 Resetting the measuring device
The entire device configuration or some of the configuration can be reset to a defined state
with the
Device reset
12.11.1 Function scope of "Device reset" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
To delivery settings
Every parameter for which a customer-specific default setting was ordered is reset
to the customer-specific value. All other parameters are reset to the factory setting.
Restart device
The restart resets every parameter with data stored in volatile memory (RAM) to
the factory setting (e.g. measured value data). The device configuration remains
unchanged.
Restore S-DAT backup
Restores the data that is saved on the S-DAT. Additional information: This function
can be used to resolve the memory issue "083 Memory content inconsistent" or to
restore the S-DAT data when a new S-DAT has been installed.
This option is displayed only in an alarm condition.
12.12 Device information
The
Device information
submenu contains all parameters that display different
information for device identification.
Navigation
"Diagnostics" menu → Device information
‣
Device information
Device tag
Serial number
Firmware version
Device name