Commissioning
Proline Cubemass C 300
106
Hauser
10.4.12 Configuring the low flow cut off
The
Low flow cut off
wizard systematically guides the user through all the parameters
that must be set to configure low flow cut off.
Navigation
"Setup" menu → Low flow cut off
‣
Low flow cut off
Assign process variable
On value low flow cutoff
Off value low flow cutoff
Pressure shock suppression
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
entry
Factory setting
Assign process variable
–
Select process variable for low
flow cut off.
• Off
• Mass flow
• Volume flow
• Corrected volume
flow
Mass flow
On value low flow cutoff
A process variable is selected
in the
Assign process variable
Enter on value for low flow cut
off.
Positive floating-
point number
Depends on country
and nominal
diameter
Off value low flow cutoff
A process variable is selected
in the
Assign process variable
Enter off value for low flow cut
off.
0 to 100.0 %
50 %
Pressure shock suppression
A process variable is selected
in the
Assign process variable
Enter time frame for signal
suppression (= active pressure
shock suppression).
0 to 100 s
0 s
*
Visibility depends on order options or device settings