Proline Cubemass C 300
Technical data
Hauser
193
16.8 Environment
Ambient temperature
range
Temperature tables
Observe the interdependencies between the permitted ambient and fluid
temperatures when operating the device in hazardous areas.
For detailed information on the temperature tables, see the separate document
entitled "Safety Instructions" (XA) for the device.
Storage temperature
–50 to +80 °C (–58 to +176 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Relative humidity
The device is suitable for use in outdoor and indoor areas with a relative humidity of
4 to 95%.
Operating height
According to EN 61010-1
• ≤ 2
000 m (6
562 ft)
• > 2
000 m (6
562 ft) with additional overvoltage protection (e.g. Hauser HAW
Series)
Degree of protection
Transmitter
• IP66/67, Type 4X enclosure, suitable for pollution degree 4
• When the housing is open: IP20, Type 1 enclosure, suitable for pollution degree 2
• Display module: IP20, Type 1 enclosure, suitable for pollution degree 2
Optional
External WLAN antenna
IP67
Vibration- and shock-
resistance
Vibration sinusoidal, in accordance with IEC 60068-2-6
• 2 to 8.4 Hz, 3.5 mm peak
• 8.4 to 2
000 Hz, 1 g peak
Vibration broad-band random, according to IEC 60068-2-64
• 10 to 200 Hz, 0.003 g
2
/Hz
• 200 to 2
000 Hz, 0.001 g
2
/Hz
• Total: 1.54 g rms
Shock half-sine, according to IEC 60068-2-27
6 ms 30 g
Rough handling shocks, according to IEC 60068-2-31
Interior cleaning
• Cleaning in place (CIP)
• Sterilization in place (SIP)