Proline Cubemass C 300
Diagnostics and troubleshooting
Hauser
161
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
444
Current input 1 to n
faulty
1. Check process
2. Check current input settings
S
Warning
453
Flow override active
Deactivate flow override
C
Warning
484
Failure mode simulation
active
Deactivate simulation
C
Alarm
485
Process variable
simulation active
Deactivate simulation
C
Warning
486
Current input simulation
active
Deactivate simulation
C
Warning
491
Current output 1 to n
simulation active
Deactivate simulation
C
Warning
492
Frequency output
simulation active
Deactivate simulation frequency output C
Warning
493
Pulse output simulation
active
Deactivate simulation pulse output
C
Warning
494
Switch output
simulation active
Deactivate simulation switch output
C
Warning
495
Diagnostic event
simulation active
Deactivate simulation
C
Warning
496
Status input simulation
active
Deactivate simulation status input
C
Warning
502
CT activation/
deactivation failed
Follow the sequence of the custody
transfer activation/deactivation: First
authorized user login, then set the DIP
switch on the main electonic module
C
Warning
520
I/O 1 to n hardware
configuration invalid
1. Check I/O hardware configuration
2. Replace wrong I/O module
3. Plug the module of double pulse
output on correct slot
F
Alarm
528
Concentration
calculation not possible
Out of valid range of the selected
calculation algorithm
1. Check concentration settings
2. Check measured values, e.g. density
or temperature
S
Alarm
529
Concentration
calculation not accurate
Out of valid range of the selected
calculation algorithm
1. Check concentration settings
2. Check measured values, e.g. density
or temperature
S
Warning
537
Configuration
1. Check IP addresses in network
2. Change IP address
F
Warning
540
Custody transfer mode
failed
1. Power off device and toggle DIP
switch
2. Deactivate custody transfer mode
3. Reactivate custody transfer mode
4. Check electronic components
F
Alarm
543
Double pulse output
1. Check process
2. Check pulse output settings
S
Warning
593
Double pulse output
simulation
Deactivate simulation pulse output
C
Warning
594
Relay output simulation Deactivate simulation switch output
C
Warning
Содержание Proline Cubemass C 300
Страница 215: ......
Страница 216: ...www addresses endress com 71574975 71574975...