Commissioning
Proline Cubemass C 300
86
Hauser
Parameter
Prerequisite
Description
User interface /
Selection / User
entry
Factory setting
Current span
–
Select current range for
process value output and
upper/lower level for alarm
signal.
• 4...20 mA (4...
20.5 mA)
• 4...20 mA NE
(3.8...20.5 mA)
• 4...20 mA US
(3.9...20.8 mA)
• 0...20 mA (0...
20.5 mA)
Country-specific:
• 4...20 mA NE
(3.8...20.5 mA)
• 4...20 mA US
(3.9...20.8 mA)
Failure mode
–
Define input behavior in alarm
condition.
• Alarm
• Last valid value
• Defined value
Alarm
Failure value
In the
Failure mode
parameter, the
Defined value
option is selected.
Enter value to be used by the
device if input value from
external device is missing.
Signed floating-point
number
0
*
Visibility depends on order options or device settings
10.4.6 Configuring the status input
The
Status input
submenu guides the user systematically through all the parameters that
have to be set for configuring the status input.
Navigation
"Setup" menu → Status input 1 to n
‣
Status input 1 to n
Assign status input
Terminal number
Active level
Terminal number
Response time status input
Terminal number