Proline Cubemass C 300
Commissioning
Hauser
85
Parameter
Description
User interface / Selection /
User entry
Factory setting
Apply I/O configuration
Apply parameterization of the freely
configurable I/O module.
• No
• Yes
No
I/O alteration code
Enter the code in order to change the I/O
configuration.
Positive integer
0
*
Visibility depends on order options or device settings
10.4.5 Configuring the current input
The
"Current input" wizard
guides the user systematically through all the parameters that
have to be set for configuring the current input.
Navigation
"Setup" menu → Current input
‣
Current input 1 to n
Terminal number
Signal mode
0/4 mA value
20 mA value
Current span
Failure mode
Failure value
Parameter overview with brief description
Parameter
Prerequisite
Description
User interface /
Selection / User
entry
Factory setting
Terminal number
–
Shows the terminal numbers
used by the current input
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
–
Signal mode
The measuring device is
not
approved for use in the
hazardous area with type of
protection Ex-i.
Select the signal mode for the
current input.
• Passive
• Active
Active
0/4 mA value
–
Enter 4 mA value.
Signed floating-point
number
0
20 mA value
–
Enter 20 mA value.
Signed floating-point
number
Depends on country
and nominal
diameter