Proline Cubemass C 300
Commissioning
Hauser
123
Navigation
"Setup" menu → Advanced setup → WLAN settings
‣
WLAN settings
WLAN IP address
Security type
WLAN passphrase
Assign SSID name
SSID name
Apply changes
Parameter overview with brief description
Parameter
Prerequisite
Description
User entry /
Selection
Factory setting
WLAN IP address
–
Enter IP address of the WLAN
interface of the device.
4 octet: 0 to 255 (in
the particular octet)
192.168.1.212
Network security
–
Select the security type of the
WLAN network.
• Unsecured
• WPA2-PSK
• EAP-PEAP with
MSCHAPv2
• EAP-PEAP
MSCHAPv2 no
server authentic.
• EAP-TLS
WPA2-PSK
WLAN passphrase
The
WPA2-PSK
option is
selected in the
Security type
parameter.
Enter the network key (8 to 32
characters).
The network key
supplied with the device
should be changed
during commissioning
for security reasons.
8 to 32-digit
character string
comprising numbers,
letters and special
characters (without
spaces)
Serial number of the
measuring device
(e.g. L100A802000)
Assign SSID name
–
Select which name will be used
for SSID: device tag or user-
defined name.
• Device tag
• User-defined
User-defined
SSID name
• The
User-defined
option is
selected in the
Assign SSID
name
parameter.
• The
WLAN access point
option is selected in the
WLAN mode
parameter.
Enter the user-defined SSID
name (max. 32 characters).
The user-defined SSID
name may only be
assigned once. If the SSID
name is assigned more
than once, the devices
can interfere with one
another.
Max. 32-digit
character string
comprising numbers,
letters and special
characters
EH_device
designation_last 7
digits of the serial
number (e.g.
EH_Cubemass_300_
A802000)
Apply changes
–
Use changed WLAN settings.
• Cancel
• Ok
Cancel
*
Visibility depends on order options or device settings