Proline Cubemass C 300
Commissioning
Hauser
97
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Switch output function
The
Switch
option is selected
in the
Operating mode
parameter.
Select function for switch
output.
• Off
• On
• Diagnostic
behavior
• Limit
• Flow direction
check
• Status
Off
Assign diagnostic behavior
• In the
Operating mode
parameter, the
Switch
option is selected.
• In the
Switch output
function
parameter, the
Diagnostic behavior
option
is selected.
Select diagnostic behavior for
switch output.
• Alarm
• Alarm or warning
• Warning
Alarm
Assign limit
• In the
Operating mode
parameter, the
Switch
option is selected.
• In the
Switch output
function
parameter, the
Limit
option is selected.
Select process variable for limit
function.
• Mass flow
• Volume flow
• Corrected volume
flow
• Target mass flow
• Carrier mass flow
• Target volume
flow
• Carrier volume
flow
• Target corrected
volume flow
• Carrier corrected
volume flow
• Density
• Reference density
• Concentration
• Temperature
• Totalizer 1
• Totalizer 2
• Totalizer 3
• Oscillation
damping
• Pressure
• Application
specific output 0
• Application
specific output 1
• Inhomogeneous
medium index
• Suspended bubbles
index
Volume flow
Assign flow direction check
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Flow direction check
option is selected in the
Switch output function
parameter.
Select process variable for flow
direction monitoring.
• Off
• Volume flow
• Mass flow
• Corrected volume
flow
Mass flow
Assign status
• The
Switch
option is
selected in the
Operating
mode
parameter.
• The
Status
option is
selected in the
Switch
output function
parameter.
Select device status for switch
output.
• Partially filled pipe
detection
• Low flow cut off
Partially filled pipe
detection
Содержание Proline Cubemass C 300