Proline Cubemass C 300
Commissioning
Hauser
95
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Maximum frequency value
The
Frequency
option is
selected in the
Operating
mode
and a process variable is
selected in the
Assign
frequency output
parameter
Enter maximum frequency.
0.0 to 10
000.0 Hz
10
000.0 Hz
Measuring value at minimum
frequency
The
Frequency
option is
selected in the
Operating
mode
and a process variable is
selected in the
Assign
frequency output
parameter
Enter measured value for
minimum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Measuring value at maximum
frequency
The
Frequency
option is
selected in the
Operating
mode
and a process variable is
selected in the
Assign
frequency output
parameter
Enter measured value for
maximum frequency.
Signed floating-point
number
Depends on country
and nominal
diameter
Failure mode
The
Frequency
option is
selected in the
Operating
mode
and a process variable is
selected in the
Assign
frequency output
parameter
Define output behavior in
alarm condition.
• Actual value
• Defined value
• 0 Hz
0 Hz
Failure frequency
In the
Operating mode
Frequency
option is selected,
in the
Assign frequency
output
process variable is selected,
and in the
Failure mode
parameter, the
Defined value
option is selected.
Enter frequency output value
in alarm condition.
0.0 to 12
500.0 Hz
0.0 Hz
Invert output signal
–
Invert the output signal.
• No
• Yes
No
*
Visibility depends on order options or device settings