Proline Cubemass C 300
Commissioning
Hauser
125
Function scope of the "Configuration management" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
Execute backup
A backup copy of the current device configuration is saved from the HistoROM
backup to the memory of the device. The backup copy includes the transmitter data
of the device.
Restore
The last backup copy of the device configuration is restored from the device
memory to the device's HistoROM backup. The backup copy includes the
transmitter data of the device.
Compare
The device configuration saved in the device memory is compared with the current
device configuration of the HistoROM backup.
Clear backup data
The backup copy of the device configuration is deleted from the memory of the
device.
HistoROM backup
A HistoROM is a "non-volatile" device memory in the form of an EEPROM.
While this action is in progress, the configuration cannot be edited via the local
display and a message on the processing status appears on the display.
10.5.8 Using parameters for device administration
The
Administration
submenu systematically guides the user through all the parameters
that can be used for device administration purposes.
Navigation
"Setup" menu → Advanced setup → Administration
‣
Administration
‣
Define access code
‣
Reset access code
Device reset
Using the parameter to define the access code
Complete this wizard to specify an access code for the Maintenance role.
Navigation
"Setup" menu → Advanced setup → Administration → Define access code
‣
Define access code
Define access code
Confirm access code