Function
Options
Info
Set plotting window
range
Here you can specify the start and end times of
the logbook entries to be displayed graphically.
Plotter begin
Selection
• First entry
• Date/Time
Factory setting
First entry
•
First entry
: Defines the first entry saved in
the logbook as the start time.
•
Date/Time
: Defines the set date/time as the
start time.
Plotter end
Selection
• Last entry
• Date/Time
Factory setting
Last entry
•
First entry
: Defines the last entry saved in the
logbook as the end time.
•
Date/Time
: Defines the set date/time as the
end time.
Show plot
Graphic display of the
logbook entries
The entries are displayed according to your
settings in the
General settings
/
Logbooks
.
Delete all entries
Action
You can delete all the absorption data logbook
entries here.
Data logbooks
Chronological list of the data logbook entries for
sensors
Data logbook 1 ... 8
<Logbook name>
This submenu is available for each data logbook
that you have set up and activated.
Source of data
Read only
Input or mathematical function is displayed
Measured value
Read only
Measured value being recorded is displayed
Log time left
Read only
Display of days, hours and minutes until logbook
is full.
‣
Pay attention to the information on
selecting the memory type in the
General
settings
/
Logbooks
menu.
Show
Events are displayed
Select a particular event to display more detailed
information.
Go to date
User entry
• Go to date
• Time
Use this function to go directly to a specific time
in the list. In this way, you avoid having to scroll
through all the information. The complete list is
always visible, however.
Show plot
Graphic display of the
logbook entries
The entries are displayed according to your
settings in the
General settings
/
Logbooks
.
Select 2nd plot
Select another data
logbook
You can view a second logbook at the same time
as the current one.
Delete all entries
Action
You can use this to delete all data logbook
entries.
Save logbooks
File format
Selection
• CSV
• FDM
‣
Save the logbook in the preferred file
format.
You can then open the saved CSV file on the PC
in MS Excel, for example, and make further edits
here.
. You can import the FDM files into
FieldCare and archive them so that they are
tamper-proof.