Operation
Liquiline System CA80NO
58
Hauser
Menu/Setup/General settings/Logbooks
Function
Options
Info
Analyzer event logbook
Selection
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is 80 % full, the device displays a
diagnostic message.
If the memory is full, there is an overflow, i. e. no
new values can be saved. The controller displays
a corresponding diagnostic message. The
memory then has to be cleared manually.
Overflow warnings
Analyzer calibration logbook
=
Fill up buffer
Calibration logbook
Selection
• Off
• On
Factory setting
Off
‣
Decide whether you want to receive a
diagnostic message if the fill buffer of the
relevant logbook overflows.
Diagnostic logbook
Configuration logbook
Analyzer data logbooks
For analyzer measured data
The entry is made automatically on completion
of the measurement. A setting does not need to
be made. The logbook is activated automatically.
SP1 is assigned to data logbook SP1.
Data logbook SP1
Assignment to measuring channel
Source of data
Read only
The assigned measuring channel is displayed
Measuring parameter
Read only
Plain-text information on the parameter that is
being recorded
Unit
Read only
Information about the unit in which the data are
available
Logbook name
Customized text, 16
characters
Line plotter
Menu to define the graphic display
Axes
Selection
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (
On
) or not
(
Off
)?
Orientation
Selection
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves should
be displayed from left to right (
Horizontal
) or
from top to bottom (
Vertical
). If you want to
display two data logbooks simultaneously, make
sure that both logbooks have the same settings
here.
X-Description
Selection
• Off
• On
Factory setting
On
‣
Decide whether a description should be
displayed for the axes and whether gridlines
should be shown. In addition, you can also
decide whether pitches should be displayed.
Y-Description
Grids
Pitches
X Pitch/Grid
distance
10 to 50%
Factory setting
10 %
‣
Determine the pitch.
Y Pitch/Grid
distance