Liquiline System CA80NO
Operation
Hauser
77
10.6 Binary inputs and outputs
Hardware options, such as the "DIO" module with 2 digital inputs and 2 digital outputs or
fieldbus module "485" enable the following:
• Via a digital input signal
• measuring range switching for conductivity (upgrade code required, )
• switching between different calibration datasets in the case of optical sensors
• external hold (for sensors)
• a cleaning interval to be triggered (for sensors)
• measurements to be started, interruption of measuring intervals
• PID controller switch-on/switch-off, e.g. via the proximity switch of the CCA250
• the use of the input as an "analog input" for pulse-frequency modulation (PFM)
• Via a digital output signal
• diagnostic states, point level switches, "Active measurement" system state, "Sample
required" information or similar states transmitted statically (similar to a relay)
• the dynamic transmission (comparable to a non-wearing "analog output") of PFM
signals, e.g. to control dosing pumps
10.6.1 Application examples
Chlorine regulation with feedforward control
CCA250
CA80
Flow, e.g. Promag53
A
BU
BK
BN
CCA250
INS
47
48
91
92
CA80
DIO
1
2
3
Dig.in
CA80: DIO
Dig.in: PFM
91
92
+
–
45
46
CA80
DIO
Dig.out: PFM
A0029239
70
Example of chlorine control with feedforward control
1
Connection of the inductive proximity switch INS of CCA250 to the digital input of the DIO module
2
Connection of the signal from a flowmeter to the digital input of the DIO module
3
Activation of a (pulse) dosing pump via the digital output of the DIO module
A
Dosing pump