Installation
Liquiline System CA80NO
18
Hauser
‣
Ensure that the unit can drain freely, without any siphoning effects.
‣
Make sure air can circulate freely at the front of the housing.
‣
Open analyzers (i.e. analyzers that are supplied without a door) may only be erected in
closed areas or in a protective cabinet or similar facility.
5.1.3
Spacing requirements when mounting
Spacing required for installing analyzer
≥330(12.99)
≥
100 (3.94)
A0036774
17
Minimum spacing required for mounting. Engineering unit
mm (in).
≤116 °
A0036775
18
Maximum opening angle
Spacing required for installing wall-mount version
750 (29.53)
120 (4.72)
Ø 11 (0.43)
Ø 10.7 (0.42)
A0036779
19
Holder unit dimensions. Engineering unit mm (in)
5.2
Mounting the analyzer on a wall
L
CAUTION
Incorrect installation can cause injury and damage the device
‣
If mounting on a wall, check that the analyzer is fully hooked into the wall holder unit
at the top and bottom, and secure the analyzer to the upper wall holder unit using the
securing screw.