Maintenance
Liquiline System CA80NO
144
Hauser
12.5 Replacing filter mats
You require the following parts:
Filter mats (part of the CAV800 maintenance kit)
1. Open and remove the fan screen on the right and left underside of the analyzer.
2. Remove the used filter mats and replace them with new ones from the maintenance
kit.
3. Fit the fan screens back on.
4.
Menu
/
Operation
/
Operating time filter mats
/
Reset
Select .
12.6 Replacing dispensers
You require the following parts:
• From the CAV800 maintenance kit:
• NORPRENE hose, ID 1.6 mm
• 10 ml dispensers with adapter
• 2.5 ml dispenser with adapter
• 1 pair of gloves that are resistant to the reagents used
Preparing to replace dispensers
1. Rinse the system.→ 142
2. Remove the bottles including the reagent hoses and the bottle tray.
3. Remove the cover of the carrier board.
Removing dispensers
1. Select
Menu
/
Operation
/
Dispenser replacement
/
Dispenser selection
.
2. Select the dispensers that you want to replace.
3. Select
Draw up dispenser
.
4. Open the dispenser holder by pressing both snap-fit hooks together and removing it.
5. Turn the dispenser counterclockwise to remove it from the Liquid Manager.
6. Remove the adapter and the dispenser from the dispenser drive. To do so, hold the
dispenser by the black block at the lower end and pull the block from the metal pin.
Inserting new dispensers
1. Screw a new dispenser onto the Liquid Manager. Make sure that the dispenser is
connected correctly.
2. Slide the dispenser holder over the snap-fit hooks. Ensure that the holder snaps into
place. Make sure that the drain hose D6 is routed behind dispenser 7.
3. Secure the cover and place the bottle tray back into the housing.
4. Connect the reagent hoses to the corresponding reagent valves.
5. Select
Menu
/
Operation
/
Dispenser replacement
/
Dispenser selection
.
6. Select the dispensers that you have replaced and press
OK
to confirm.
7. Select
Reset operating time counter
.
8. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
9. Select all the bottles and press
OK
to confirm.
10. Select
Confirm bottles inserted
.