Operation
Liquiline System CA80NO
68
Hauser
3. Enter the new password again.
The lock password has been changed successfully.
Press the navigator for longer to return to the measuring mode.
10.3 Analyzer
Menu/Setup/Analyzer
Function
Options
Info
Mode
Read only
Manual, automatic or fieldbus
Device tag
Customized text,
32 characters
Factory setting
Analyzer_serial number
‣
Select any name for your analyzer. Use the
TAG name for example.
Meas. param
Read only
Measuring range
Read only
10.3.1 Advanced setup
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
Measuring value
Main value
Selection
• NO2
• NO2-N
Factory setting
NO2-N
If another main value is selected, this affects
other display screens. However, if another main
value is selected, these display screens only
change by a constant factor.
Unit
Selection
• mg/l
• µg/l
• ppm
• ppb
Factory setting
mg/l
or
µg/l
Main value format
Selection
• #
• #.#
• #.##
• Auto
Factory setting
#.#
Measuring range
Options depend on the
parameter
Only with optional dilution function
When selecting the measuring range, note that
the dilution function is only available for a
certain measuring range.
Internal dilution factor
1 ... 20
Factory setting
1
Specify the desired dilution factor.