Operation
Liquiline System CA80NO
62
Hauser
Menu/Setup/(General settings or Inputs<Sensor channel>)/Extended setup/Diagnostics settings/Diag.
behavior
Function
Options
Info
Diag. output
Selection
• None
• Alarm relay
• Binary output
• Relay 1 to n (depends on
the device version)
Factory setting
None
Before being able to assign the message to an
output you must first configure a relay output to
Diagnostics
.
(
Menu
/
Setup
/
Outputs
: Assign the
Diagnostics
function and set the
Operating mode
to
as
assigned
.)
An alarm relay is always available, regardless of the device version. Other relays are optional.
Cleaning program
(for sensors)
Selection
• None
• Cleaning 1
• Cleaning 2
• Cleaning 3
• Cleaning 4
Factory setting
None
‣
Decide whether the diagnostic message
should trigger a cleaning program.
You can define the cleaning programs under:
Menu
/
Setup
/
Additional functions
/
Cleaning
.
Detail information
Read only
Here you can find more information on the
diagnostic message and instructions on how to
resolve the problem.
PROFIBUS DP
Menu/Setup/General settings/Extended setup/PROFIBUS
Function
Options
Info
Enable
Selection
• Off
• On
Factory setting
On
You can switch off communication at this point.
The software can then only be accessed via local
operation.
Termination
Read only
If the device is the last in the bus, you can
terminate via the hardware.
Bus address
1 to 125
If you have addressed the bus via hardware (DIP
switches on the module, ), you can only read the
address here.
If an invalid address is set via the hardware, you
have to assign a valid address for your device
either here or via the bus.
Ident number
Selection
• Automatic
• PA-Profile 3.02 (9760)
• Manufacturer specific
Factory setting
Automatic