Liquiline System CA80NO
Maintenance
Hauser
143
Replacing bottles
1. Replace the removed bottles with fresh reagent .
Reagents must be prepared according to the instructions for mixing a reagent.
2. Reconnect the hoses to the Liquid Manager.
3.
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle
selection
must be selected.
4. Select all the bottles that you have replaced and press
OK
to confirm.
5.
Confirm bottles inserted
must be selected.
6. If bottle level monitoring is enabled (
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Diagnostics settings
/
Bottles
), it can be reset under
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Reset filling levels
.
7. Zero point calibration of the system must be performed after a replacement routine .
8. After the zero point calibration, select
Menu
/
Operation
/
Manual operation
/
Determine calibration factor
.
9. Go back to
MODE
/
Continue automatic mode
or
MODE
/
Start automatic mode
to
start normal measurement.
12.4 Replacing the hoses
You require the following parts:
• From the CAV800 maintenance kit:
• NORPRENE hose, ID 1.6 mm
• C
‐
Flex hose, ID 3.2 mm
• C
‐
Flex hose, ID 6.4 mm
• Hose connector
• 1 pair of gloves that are resistant to the reagents used
1. Press the
MODE
soft key and select the
Manual mode
.
The display shows
Current mode
-
Manual
.
Wait until all operations have stopped. Ongoing operations can be stopped under
Menu
/
Operation
/
Manual operation
.
There should not be any more sample left in the sample collecting vessel.
3. Remove the bottles and the bottle tray.
4. Remove the cover of the carrier board.
5. Replace with hoses of the same diameter and length. Attach hose labels to the new
hoses.
6. Route outlet hose D6 behind dispenser 7.
7. Secure the cover and place the bottle tray back into the housing.
8. Connect the bottles to the appropriate covers and hoses.
9. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
10. Select all the bottles and press
OK
to confirm.
11. Select
Confirm bottles inserted
.
12. Press the
MODE
soft key and select
Continue automatic mode
or
Start automatic
mode
.