Liquiline System CA80NO
Operation
Hauser
67
5. Use
OK
to confirm or cancel and assign a new file name.
Your setup is saved on the SD card in the "Device" folder.
You cannot upload the exported setup to the device again. You must use the
Save
setup
. This is the only way you can save a setup to an SD card and reload it later on or
upload it to other devices.
Activation code
You require activation codes for:
• Additional functions, e.g. fieldbus communication
• Firmware upgrades
If your original device has activation codes, these can be found on the nameplate. The
corresponding device functions are activated at the factory. You only require the codes
if servicing the device.
1. Enter activation code:
Menu
/
Setup
/
General settings
/
Extended setup
/
Data
management
/
Activation code
.
2. Confirm your entry.
The new hardware or software function is activated and can be configured.
Functions that are enabled by an activation code:
Function
Activation code beginning with
Two current outputs (BASE2-E module only)
081...
Web server
351...
HART
0B1...
PROFIBUS DP
0B3...
Modbus TCP
0B8...
Modbus RS485
0B5...
EtherNet/IP
0B9...
PROFINET
0B7...
Cooling
0F1...
Feedforward control
220...
Chemoclean Plus
25...
Collecting vessel
20...
Measuring channels
28...
Configure the mathematical function
Cation exchanger capacity
301...
1)
Via Ethernet socket of the Base2 module, for versions without an Ethernet fieldbus
2)
Availability depends on measuring parameter
Changing the password
You can lock the operating keys with a password (access the context menu by pressing the
navigator for longer). The keys can then only be enabled by entering the right password.
You can set the password to lock the keys here:
Menu
/
Setup
/
General settings
/
Extended
setup
/
Data management
/
Change lock password
.
1. Enter the current password (factory setting 0000).
Enter new lock password
2. Enter a new password.
Confirm new lock password