Liquiline System CA80NO
Electrical connection
Hauser
39
LEDs on front of module
LED
Designation
Color
Description
RJ45
LNK/ACT
GN
• Off = Connection is not active
• On = Connection is active
• Flashing = Data transmission
RJ45
10/100
YE
• Off = Transmission rate 10 MBit/s
• On = Transmission rate 100 MBit/s
PWR
Power
GN
Supply voltage is applied and module is initialized
BF
Bus failure
RD
Not used
SF
System failure
RD
Device error
COM
Communication
YE
Modbus message sent or received
DIP switches on front of module
DIP
Factory
setting
Assignment
1-128
ON
Bus address (→ "Commissioning/communication")
OFF
Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service
OFF
If the switch is set to
"ON"
, the user settings for Ethernet addressing are saved
and connection settings programmed into the device at the factory are activated:
IP address=192.168.1.212, Subnet mask=255.255.255.0, Gateway=0.0.0.0,
DHCP=Off.
If the switch is set to
"OFF"
, the saved user settings are reactivated.
6.5
Hardware settings
6.5.1
Bus termination (module 485 only)
There are 2 ways to terminate the bus:
1. Internal termination
(via DIP switch on module board)
"ON"
"OFF"
56
DIP switch for internal termination
‣
Using a suitable tool such as a tweezer, move all four DIP switches to the "ON" position.
The internal termination is used.