Liquiline System CA80NO
Operation
Hauser
61
• Logbook name
Assign a name. Example: "01".
• Source of data
Select a data source. Example: Signal of binary input 1.
• Measured value
Select the measured value to be recorded. Example: pH value.
• Scan time
Specify the time interval between two logbook entries.
• Data logbook
Activate the logbook: specify the data storage method.
2. ../
Finished
: Perform the action.
The device shows the new logbook in the list of data logbooks.
3. Select data logbook "01".
Additional display:
Log time left
.
4. Only in the case of
Fill up buffer
:
Decide to set
Overflow warning
:
On
or
Off
.
On
: The device displays a diagnostic message in the event of memory overflow.
5.
Line plotter
submenu: Specify the type of graphic representation.
10.2.5 Advanced setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/(General settings or Inputs<Sensor channel>)/Extended setup/Diagnostics settings/Diag.
behavior
Function
Options
Info
List of diagnostic messages
‣
Select the message to be changed. Only then
can you make the settings for this message.
Diag. code
Read only
Diagnostic message
Selection
• On
• Off
Factory setting
Depends on the message
You can deactivate or reactivate a diagnostic
message here.
Deactivating means:
• No error message in the measuring mode
• No error current at the current output
Failure current
Selection
• On
• Off
Factory setting
Depends on the message
‣
Decide whether an error current should be
output at the current output if the diagnostic
message display is activated.
In the event of general device errors, the
error current is output at all the current
outputs. In the event of channel-specific
errors, the error current is only output at
the assigned current output.
Status signal
Selection
• Maintenance (M)
• Out of specification (S)
• Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error
categories in accordance with NAMUR NE 107.
‣
Decide whether you want to change a status
signal assignment for your application.