Liquiline System CA80NO
Operation
Hauser
85
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Source of data
Selection
• None
• Sensor inputs
• Binary inputs
• Controller
• Fieldbus signals
• Mathematical functions
Factory setting
None
Source, whose value should be read out as a
frequency over the binary output.
Measuring value
Source of data
≠ Controller
Selection
Depends on:
Source of data
‣
Choose the measured value which should be
output as the frequency via the binary
output.
Actuator type
Source of data
= Controller
Selection
• None
• Bipolar
• U
• Unipolar-
Factory setting
None
‣
Determines which controller component the
connected actuator, e. g. the dosing pump,
should trigger.
Bipolar
"Split range"
U
Part of the manipulated variable that the
controller uses to increase the process value
Unipolar-
For connected actuators that decrease the
controlled variable
Hold behavior
Selection
• Freeze
• Fixed value
• None
Factory setting
None
Freeze
The device freezes the last value.
Fixed value
You define a fixed current value that is
transmitted at the output.
None
A hold does not affect this output.
Hold value
Hold behavior
=
Fixed value
0 to 100 %
Factory setting
0 %
Error behavior
Selection
• Freeze
• Fixed value
Factory setting
Fixed value
Freeze
The device freezes the last value.
Fixed value
You define a fixed current value that is
transmitted at the output.
Error value
Error behavior
=
Fixed value
0 to 100 %
Factory setting
0 %
1)
x:y = slot no. : input number
10.7 Signal outputs
10.7.1 Current outputs
Liquiline System CA80 has two analog current outputs by default.
You can configure additional current outputs with extension modules.
Setting the current output range
‣
Menu
/
Setup
/
General settings
: Select
0..20 mA
or
4..20 mA
.