Liquiline System CA80NO
Diagnostics and troubleshooting
Hauser
125
No.
Message
Factory settings
Tests or remedial action
S
D
F
718
Cuvette covering
M
On
Off
Photometer check
High degree of contamination - reliable
measurement will not be possible soon
1.
Clean analyzer manually.
2.
Contact the Service Department.
719
Cuvette covering
F
On
Off
Photometer check buildup, high degree of
contamination - measurement no longer possible
1.
Start cleaning.
2.
Contact the Service Department.
726
Liquids warning
M
On
Off
Consumable liquids, warning
Measurement can still take place.
Possible reasons
• The level of one or more liquids is low
• One or more liquids almost past their shelf life.
‣
Top up/replace the liquids in question and
reset the counter in
Diagnostics
/
Operating
time information
.
727
Liquids alarm
F
On
Off
Consumable liquids, alarm
Measurement can still take place.
Possible reasons
• The level of one or more liquids is low
• One or more liquids past their shelf life.
‣
Top up/replace the liquids in question and
reset the counter in
Diagnostics
/
Operating
time information
.
729
Filter candle change
M
On
Off
• Filter cartridge needs replacement
• The limit value for operating hours has been
exceeded
‣
Replace the filter cartridge for the sample
preparation system and reset the operating
hours counter in the Diagnostics menu.
730
Cleaning solution
M
On
Off
• Level warning for cleaner of sample preparation
system
• Depending on the duration of cleaning, the
cleaning interval and external events, the
remaining quantity is sufficient for a few hours
or days
1.
Top up cleaner for the sample preparation
system.
2.
Check the level switch for the cleaning
solution.
731
Leakage detected
F
On
Off
Leak detected in the sample preparation system or
analyzer
1.
Check hoses and connections.
2.
Check solenoid valves.
3.
Check leak sensor.
4.
Check the free outlet on the analyzer.
5.
Replace defective parts and, if necessary,
reset the operating hours counters in the
Diagnostics menu.
732
Wear parts alarm
F
On
On
One or more wear parts has exceeded its service
life.
‣
Replace the wear parts in question and reset
the counter in
Diagnostics
/
Operating time
information
.