Operation
Liquiline System CA80NO
74
Hauser
Menu/Setup/Sample preparation
Function
Options
Info
Compressed air
(CAT820)
Selection
• Available
• Not available
Factory setting
Depends on the device
version
Settings for sample preparation with or
without compressed air cleaning.
In the event of retrofitting, the compressed air
can be activated here.
Cleaning mode
Selection
• On
• Off
Factory setting
On
Activates or deactivates automatic
compressed air cleaning of the pump-to-filter
hose and of the filter.
Cleaning interval
Selection
0:30 to 4:00 HH:MM
Factory setting
2:00
Cleaning interval of the automatic compressed
air cleaning system
Cleaning duration
Selection
10 to 60 s
Factory setting
30 s
Cleaning duration of the automatic
compressed air cleaning system
Heatings
(CAT820, CAT860)
Cabinet
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Hose filter
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Hose analyzer
Selection
• Available
• Not available
Factory setting
Depends on the device
version
In the event of retrofitting, the heating can be
activated here.
Diagnostics settings
Limit change filter
(CAT820, CAT860)
Function
Selection
• On
• Off
Factory setting
Off
Diag. code 729
Read only
Warning limit
Selection
01
‐
00 to 99
‐
00 (DD
‐
HH)
Factory setting
60-00
Limit change hose
(CAT820, CAT860)