Diagnostics and troubleshooting
Liquiline System CA80NO
128
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
992
pH calculation range
F
On
On
pH calculation outside the measuring range
993
rH calculation range
F
On
On
rH calculation outside the measuring range
994
Difference conduct.
F
On
On
Dual conductivity outside the measuring range
1)
Status signal
2)
Diagnostic message
3)
Failure current
11.6.2 Sensor-specific diagnostic messages
Operating Instructions "Memosens", BA01245C
11.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status. Furthermore,
various service functions are available.
The following messages are directly displayed every time you enter the menu:
• Most important message
Diagnostic message recorded with the highest criticality level
• Past message
Diagnostic message whose cause is no longer present.
All the other functions in the Diagnostics menu are described in the following chapters.
11.8 Diagnostics list
All the current diagnostic messages are listed here.
A time stamp is available for each message. Furthermore, the user also sees the
configuration and description of the message as saved in
Menu
/
Setup
/
General settings
/
Extended setup
/
Diagnostics settings
/
Diag. behavior
.
11.9 Logbooks
11.9.1 Available logbooks
Types of logbooks
• Logbooks physically available (all apart from the overall logbook)
• Database view of all logbooks (=overall logbook)
Logbook
Visible in
Max.
entries
Can be
disabled
Logbook
can be
deleted
Entries
can be
deleted
Can be
exported
Overall logbook
All events
20000
Yes
No
Yes
No
Calibration logbook
Calibration events
75
(Yes)
No
Yes
Yes
Operation logbook
Configuration events
250
(Yes)
No
Yes
Yes
Diagnostics logbook
Diagnostic events
250
(Yes)
No
Yes
Yes
Analyzer event
logbook
Analyzer events
19500
No
No
Yes
Yes
Analyzer calibration
logbook
Analyzer calibration
logbook
250
(Yes)
No
Yes
Yes