Operation
Liquiline System CA80NO
72
Hauser
Menu/Setup/Analyzer/Calibration
Function
Options
Info
Date
01.01.1970 to 07.02.2106
Factory setting
DD.MM.YYYY
Time
00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24 h)
Calibration interval
0-01 to 90-00 (DD-HH)
Factory setting
02-00
For configuring the time interval of the
calibration/adjustment.
Next calibration
Mode
=
Automatic
Read only
Zero point
Read only
Calibration factor
Read only
Relationship of the measured concentration to
the pre-specified concentration of the calibration
standard.
Settings
Nominal concentration
Factory setting
For configuring the concentration of the
calibration standard solution.
Depends on the settings under
Menu
/
Setup
/
Analyzer
/
Extended setup
/
Measuring value
/
Main value
Automatic cleaning
Selection
• Off
• On
Factory setting
On
For specifying whether cleaning will take place
before every calibration/adjustment (only in the
automatic mode).
10.4 Sample preparation
The menu displayed depends on the connected sample preparation system. The menu
item does not appear if a Liquiline System CA80 analyzer without a collecting vessel is
connected. The Liquiline System CAT860 can only be operated with a Liquiline
System CA80 single-channel device.
Menu/Setup/Sample preparation
Function
Options
Info
Installation type
Selection
• Others
• Pressurized pipe
sampling
Factory setting
Others
Others:
e.g. Liquiline System CAT820 / CAT860
Pressurized pipe sampling:
Sample preparation that is installed in a
pressure pipe, e.g. Liquiline System CAT810.
If the installation type selected is
Others
: